Features and Benefits
- Operational Environment: Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
- Storage Capacity: 5.2 cu. ft. gross volume
- Door: One swing glass door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Three shelves (two adjustable/one fixed) with guard rail on back
- Leveling legs. Note: 4" of clearance on all sides must be maintained for adequate ventilation
- Interior lighting: Shielded, switched LED lighting, full coverage, balanced spectrum
- Airflow Management: Forced Air technology, patent pending
- External Probe Access: Rear wall port (3/8") diameter
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior Materials: White powder coated steel
- Access Control: Pyxis®, Omnicell® and AcuDose RX® compatible
- Product Weight: 96 lbs.
- Shipping Weight: 132 lbs.
- Rated Amperage: 1.3 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification:
- Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories
- Temperature monitor device (TMD) complies with the current CDC guidelines, with 3 years certification of calibration, “buffered†probe in the product simulated solution, min/max memory. F/C switchable, field installable, and visual & audible temp alarm
Technical Specifications
- Refrigeration System
- Compressor: Hermetic, High Performance
- Refrigerant: EPA SNAP Compliant, R600A, ISOBUTANE
- Condenser: Hybrid fin tube with low noise fan
- Evaporator: Plate Wall
- Defrost: Cycle, Optimized, Zero Energy
- Performance
- Uniformity (Cabinet Air): +/- 1.4°C
- Stability (Cabinet Air): +/- 1.3°C
- Maximum Temperature Variation (Cabinet Air): +/- 1.7°C
- Temperature Rise After an After 8 Sec Door Openings: Temperature did not exceed 6.4°C at any probe for all required NSF/ANSI 456 testing protocols
- Recovery After 3 Min Door Opening: All probes recover to under 8°C within 4.8 min.
- Energy Consumption: 1.15 KWh/day
- Average Heat Rejection: 1.57 KWh/day (224 BTU/h)
- Noise Pressure Level (dBA): 41 or less installed
- Pull Down Time to Nominal Operating Temp: 42 min
- Controller, Configuration, Alarms and Monitoring
- Controller Technology: Parametric, microprocessor, LED display with 0.1°C resolution
- Temperature Setpoint Range: 1°C to 10°C (Setpoint must remain unaltered from the factory setting to remain compliant with NSF/ANSI 456 performance requirements)
- Display Probe: Calibrated, stainless steel
- External Alarm Connection
- State switching remote alarm contacts
- Visual and audible indicators
- Alarms: High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
- Simulator Ballast: 20 ml bottle, glass bead thermal media
Warranty Information
- 2-Year Warranty on Parts and Labor, Excluding Display Probe Calibration
- 5-Year Warranty on Compressor Warranty
RETURNS :
For complete details of our return policy and process, please refer to our general return policy page. We take the responsibility to make sure your product
arrives in brand new working condition. Should your item arrive damaged, we will replace the product at no cost to you.
CANCELLATIONS: Please note that any specially ordered or custom-made products are non-cancellable once production has begun. Unfortunately, we are unable to
accept returns on such items.
AVAILABILITY NOTE: Availability of products may vary based on the manufacturer’s current stock levels.