Features and Benefits
- 3 measurement modes (Monitor, Spot-check, Triage)
- High-sensitivity 8” TFT color touch screen with self-explanatory user interface and integrated help - function
- Displays SYS, DIA, MAP, Pulse, Temp, Plethysmogram
- Clinical grade NIBP including averaging mode
- Programmable NIBP averaging mode (2-5 readings)
- Available Riester, Masimo or Nellcor* Oximax SpO2
- Available with predictive Filac 3000 Thermometer
- Connectivity with the EMR - follows the HL7 standard
- LAN Connectivity as standard
- Programmable alarms (audible, visible) with 3 priority levels
- Supports nurse call RJ11 port
- Available as wall, desk and floor model
- Internal memory for 5,000 measurements
- Lithium-ion battery with up to 17 hours ON time
- Power supply for up to 3 diagnostic handles
- Maintenance-free High Performance LED illumination
- Wide range of EENT ri-scope L instruments available (Otoscope and Ophthalmoscopes are sold separately)
- Integrated Specula dispenser
- Easy and discrete mounting
- Optional wireless connection available
- Optional integrated printer available
- Optional mobile stand available
- Riester RVS-200 Advanced Vital Signs Monitor
- (1) NIBP extension tube
- (2) NIBP cuffs (adult & adult large incl. connectors)
- (2) Integrated handles for ri-scope Otoscope and Ophthalmoscopes
- (1) SpO2 sensor adult (Riester, Masimo or Nellcor)
- (1) SpO2 extension cable (Riester, Masimo or Nellcor)
- (1) Oral temperature probe (For Optional Thermometer)
- (25) Probe cover (For Optional Thermometer)
- User Manual
Technical Specifications
- Monitor
- Power Supply: 100 – 240 V AC, 50/60 Hz
- Battery:
- Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh, Charging time to 100%: 6 hours, Operating time approx. 11 hours (new and fully charged battery at 25°C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval)
- Low battery indicator
- Dimensions:
- Length: 10.63" / 270 mm
- Width: 8.66" / 225 mm
- Height: 7.87" / 200 mm
- Alarm:
- Visible LEDs (Yellow, Red, Blue, Green, Orange)
- Audible Speaker
- Gives audible alarm, QRS tone
- Supports Pitch Tone and multi-level volume
- Alarm tones meet the requirement of IEC 60601-1-8.
- Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone.
- Ports:
- 1 x AC power inlet
- 1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
- 4 x USB ports
- 1 x Equipotential grounding point
- 1 x RJ11 connector for nurse call DC out port 15V/1.2 A
- Clinical-grade NIBP (non-invasive blood pressure)
- Measuring method: Automatic oscillometric
- Measurement range:
- Adult: SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
- Pediatric: SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
- Neonate: SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg
- Cuff pressure range: 0 – 280 mmHg
- Resolution: 1 mmHg
- Pressure Accuracy
- Static ±3 mmHg
- Clinic ±5 mmHg, standard deviation 8 mmHg
- Pulse Rate Range: 40 – 240 bpm
- Pulse Rate Accuracy: 1 bpm
- Initial Inflation Pressure: Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg Double Over-pressure protection (hardware and software) Adult 297 ± 3 mmHg, Pediatric 252 ± 3 mmHg , Neonate 147 ± 3 mmHg
- Alarm Range:
- SYS 0 – 300 mmHg
- DIA 0 – 300 mmHg
- MAP 0 – 300 mmHg
- SpO2
- Riester
- Measurement range: 0 – 100%
- Resolution: 1%
- Accuracy:
- ±2%(70 to 100%)
- ±3% (40% to 69%)
- Unspecified (0% to 39%)
- Alarm range: 0 – 100%
- Average time: 4 sec, 8 sec, 16 sec
- Pulse rate
- Measurement range: 20 – 250 bpm
- Resolution: 1 bpm
- Accuracy: ±1% or ±1 bpm whichever is greater
- Nellcor
- Measurement range: 0-100%
- Resolution: 1%
- Accuracy:
- 70 – 100%: ±2% (adult, pediatric)
- 70 – 100%: ±3% (neonate)
- 0 – 69%: unspecified
- Alarm range: 0 – 100%
- Average time: 8 sec, 16 sec
- Pulse rate
- Measurement range: 20 – 300 bpm
- Resolution: 1 bpm
- Accuracy: 20 – 250 bpm: ±3 bpm
- 251 – 300 bpm: unspecified
- Masimo
- Measurement range: 0 – 100%
- Resolution: 1%
- Accuracy
- 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
- 70 – 100%: ±3% (neonate, non-motion conditions)
- 70 – 100%: ±3% (motion conditions)
- 0 – 69%: unspecified
- Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
- Pulse Rate
- Measurement range: 25 – 240 bpm
- Resolution: 1 bpm
- Accuracy
- ±3 bpm (non-motion conditions)
- ±5 bpm (motion conditions
- Riester
- Predictive Thermometer (Optional)
- Oral ( Quick Mode )
- Non-fever Temps: 3-5 seconds
- Fever Temps: 8-10 seconds
- Oral ( Standard Mode ): 6-10 seconds
- Axillary Mode: 8-12 seconds
- Rectal Mode: 10-14 seconds
- Direct Mode (all sites): 60-120 seconds
- Measurement accuracy: 35.5ºC - 42ºC
- Direct Mode (all sites): ±0.1ºC (±0.2ºF)
- Standard Prediction Mode (all sites): ±0.1ºC (±0.2ºF)
- Quick Prediction Mode (oral only): ± 0.3ºC (±0.5º)
- Oral ( Quick Mode )
Warranty Information
- 2-Year Warranty
RETURNS :
For complete details of our return policy and process, please refer to our general return policy page. We take the responsibility to make sure your product
arrives in brand new working condition. Should your item arrive damaged, we will replace the product at no cost to you.
CANCELLATIONS: Please note that any specially ordered or custom-made products are non-cancellable once production has begun. Unfortunately, we are unable to
accept returns on such items.
AVAILABILITY NOTE: Availability of products may vary based on the manufacturer’s current stock levels.